OKLEN Fitting Guide

Orthokeratology fitting. Initial lens selection by sagitta, power code and diameter.

OKLEN is an annually replaced RGP lens for orthokeratology: a 4-curve design for compensation up to −5.00 D (OKLEN) and a 5-curve design from −5.00 D to −10.00 D (OKLEN Plus). The initial lens is selected from keratometry, topography, refraction and corneal diameter.

Selecting the Initial Lens

1

Sagitta (letter)

From the parameter table, take the sagitta for the medium keratometry (Km), then adjust by the medium corneal eccentricity (em).

2

Power code (number)

The power code matching the sphere of refraction plus ½ of the corneal astigmatism (CA).

3

Diameter

The lens diameter must encompass 95% of the corneal diameter.

Eccentricity (em) to Sagitta Adjustment

e < 0.40Tight fit — 2 steps steeper (toward ZE)
0.40 ≤ e < 0.45Tight fit — 1 step steeper (toward ZE)
0.45 ≤ e ≤ 0.55Use theoretical sagitta
0.55 < e ≤ 0.60Flat fit — 1 step flatter (toward A)
e > 0.60Flat fit — 2 steps flatter (toward A)

Insertion & Assessment

Insert with preservative-free artificial tears; wait about 15 minutes for tear stabilisation. Instil fluorescein and evaluate centration, diameter, movement, VA (expected: maximum VA) and over-refraction with a cobalt-blue + yellow filter. Remove and check corneal integrity.

Follow-up Schedule

1 day1 week1 month3 months6 months

Children & adolescents with progressing myopia: measure axial length at baseline and each review to confirm control. A loose or tight fluorogram is acceptable while comfort, visual clarity and corneal integrity are maintained.

Need fitting support?

Our technical team assists with difficult fits and confirms specifications before order.

Contact Us
VISIORA

Saudi Arabia's first specialized contact lens manufacturing facility.

sales@visiora.net Riyadh, Saudi Arabia

© 2026 VISIORA. All rights reserved.

ISO 13485 Certified

ISO 13485

Medical Devices Quality Management