Use the eye shape to select an initial OraSight trial lens. Evaluate the fit on the eye after the lens settles, then refine each zone as needed.
Select the initial trial lens
Empirical Fit
Start with an empirical fit based on the measurements and shape of the eye.
AC Depth estimate
Convert AC Depth to microns, add 850 µm and select the closest Sag in the fitting reference. Edge Sag is the trial lens specification measured at the lens edge.
AC Depth and Kmax
Use the AC Depth and Kmax chart in the fitting reference to choose a starting trial lens. Confirm the selection on the eye after settling.
Check the fit
Apply the lens and assess Clearance
Fill the lens with preservative-free saline and fluorescein. Check for bubbles, Clearance and centration under diffuse blue light. Measure Apical Clearance using a slit-lamp optic section or OCT. After settling, if Central Clearance exceeds 500 µm, try a lower Sag and let the replacement lens settle for 10–15 minutes. If Clearance is below 50 µm or the lens touches the cornea, increase Sag by one or two trial steps.
Refine the Landing Zone
Assess the Limbal Vault and look for compression or Edge Lift. Flatten the Landing Zone if there is conjunctival compression, resistance on removal or staining after removal. Steepen it if there is excessive Edge Lift, tear exchange or movement. Compare both meridians. The fitting guide suggests increasing Toricity when there is Edge Lift or tear exchange at the Steep Meridian with vessel blanching at the Flat Meridian, and decreasing it when these signs are reversed. Reassess the fit after adjustment.
Need fitting support?
Our technical team assists with difficult fits and confirms specifications before order.
Contact Us